Determination of the Distal Dose Edge in a Human Phantom by Measuring the Prompt Gamma Distribution: A Monte Carlo Study

Citations

WEB OF SCIENCE

5
Citations

SCOPUS

5

초록

The close relationship between the proton dose distribution and the distribution of prompt gammas generated by proton-induced nuclear interactions along the path of protons in a water phantom was demonstrated by means of both Monte Carlo simulations and limited experiments. In order to test the clinical applicability of the method for determining the distal dose edge in a human body, a human voxel model, constructed based on a body-composition-approximated physical phantom, was used, after which the MCNPX code was used to analyze the energy spectra. and the prompt gamma yields from the major elements composing the human voxel model; finally, the prompt gamma distribution, generated from the voxel model and measured by using an array-type prompt gamma detection system, was calculated and compared with the proton close distribution. According to the results, effective prompt gammas were produced mainly by oxygen, and the specific energy of the prompt gammas, allowing for selective measurement, was found to be 4.44 MeV. The results also show that the distal dose edge in the human phantom, despite the heterogeneous composition and the complicated shape, can be determined by measuring the prompt gamma distribution with an array-type detection system.

키워드

Proton therapyPrompt gammaDistal dose edgeHuman voxel modelMonte Carlo simulationDETECTION SYSTEMPROTON-BEAMSIMULATION
제목
Determination of the Distal Dose Edge in a Human Phantom by Measuring the Prompt Gamma Distribution: A Monte Carlo Study
저자
Min, Chul HeeLee, Han RimYeom, Yeon SuCho, SungkooKim, Chan Hyeong
DOI
10.3938/jkps.56.2059
발행일
2010-06
유형
Article; Proceedings Paper
저널명
Journal of the Korean Physical Society
56
6
페이지
2059 ~ 2062