Membrane-embedded disposable microfluidic chip for cardiac biomarker screening

Citations

SCOPUS

0

초록

We report a simple and multifunctional microfluidic cartridge with electrochemical (EC) sensing for cardiac biomarker screening including a passive valve, reagent storage and electrochemical sensors from a layer-by-layer rapid prototyping technique. By embedding multiple membranes in the device, immunoassay procedures were performed by controlling fluidic paths and releasing reagents to obtain proper reactions on the electrodes. With the disposable multifunctional cartridge device, we was able to detect as low as 0.25 ng/mL of Creatine Kinase (CK)-Myocardial Band (MB) automatically.

키워드

electrochemical sensorpassive valvereagent storagecardiac biomarker screeningElectrochemical sensorsProjectilesCardiac biomarkersCreatine kinaseLayer by layerMicrofluidic chipPassive valvesMicrofluidics
제목
Membrane-embedded disposable microfluidic chip for cardiac biomarker screening
저자
Lee, JoohyungKim, JitaeSong, Si monKim, Jungkyu
발행일
2014-10
유형
Conference Paper
저널명
18th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS 2014
페이지
1551 ~ 1553