Smartwatch feedback device for high-quality chest compressions by a single rescuer during infant cardiac arrest: a randomized, controlled simulation study

Citations

WEB OF SCIENCE

20
Citations

SCOPUS

25

초록

Objective According to the guidelines, rescuers should provide chest compressions (CC) similar to 1.5 inches (40 mm) for infants. Feedback devices could help rescuers perform CC with adequate rates (CCR) and depths (CCD). However, there is no CC feedback device for infant cardiopulmonary resuscitation (CPR). We suggest a smartwatch-based CC feedback application for infant CPR. Participants and methods We created a smartwatch-based CC feedback application. This application provides feedback on CCD and CCR by colour and text for infant CPR. To evaluate the application, 30 participants were divided randomly into two groups on the basis of whether CC was performed with or without the assistance of the smartwatch application. Both groups performed continuous CC-only CPR for 2 min on an infant mannequin placed on a firm table. We collected CC parameters from the mannequin, including the proportion of correct depth, CCR, CCD and the proportion of correct decompression depth. Results Demographics between the two groups were not significantly different. The median (interquartile range) proportion of correct depth was 99 (97-100) with feedback compared with 83 (58-97) without feedback (P = 0.002). The CCR and proportion of correct decompression depth were not significantly different between the two groups (P = 0.482 and 0.089). The CCD of the feedback group was significantly deeper than that of the control group [ feedback vs. control: 41.2 (39.8-41.7) mm vs. 38.6 (36.1-39.6) mm; P= 0.004]. Conclusion Rescuers who receive feedback of CC parameters from a smartwatch could perform adequate CC during infant CPR. European Journal of Emergency Medicine 26: 266-271 Copyright (c) 2018 The Author(s). Published by Wolters Kluwer Health, Inc.

키워드

applicationcardiopulmonary resuscitationpaediatric basic life supportsmartwatchCARDIOPULMONARY-RESUSCITATIONCPR PERFORMANCEDEPTHSMARTPHONEEPIDEMIOLOGYIMPROVEMENTGUIDELINESSUPPORT
제목
Smartwatch feedback device for high-quality chest compressions by a single rescuer during infant cardiac arrest: a randomized, controlled simulation study
저자
Lee, JuncheolSong, YeongtakOh, Jae hoonChee, YoungjoonAhn, ChiwonShin, HyungooKang, Hyung gooLim, Tae Ho
DOI
10.1097/MEJ.0000000000000537
발행일
2019-08
유형
Article
저널명
European Journal of Emergency Medicine
26
4
페이지
266 ~ 271