다중 슬릿 즉발감마선 카메라를 위한 이중모드 신호처리 모듈 개발

Development of Dual-mode Signal Processing Module for Multi-slit Prompt-gamma Camera
  • 박종훈
  • 신동호
  • 이세병
  • 이한림
  • 김찬형
  • 외 2명

초록

In proton therapy, in vivo proton beam range verification is very important to deliver conformal dose to the target volume and minimize unnecessary dose to normal tissue. For this purpose, a multi-slit prompt-gamma camera module made of 24 scintillation detectors and 24-channel signal processing system is under development. In the present study, we have developed and tested a dual-mode signal processing system, which can operate in the energy calibration mode and the fast data acquisition mode, to process the signals from the 24 scintillation detectors. As a result of performance test, using the energy calibration mode, we were able to perform energy calibration for the 24 scintillation detectors at the same time and determine the discrimination levels for the detector channels. Further, using the fast data acquisition mode, we were able to measure a prompt-gamma distribution induced by a 45 MeV proton beam. The measured prompt gamma distribution was found similar to the proton dose distribution at the distal fall-off region, and the estimated beam range

키워드

Proton therapyMulti-slit prompt-gamma cameraMulti-slit collimatorScintillation detector arrayDual-mode signal processing module양성자 치료다중 슬릿 즉발감마선 카메라다중 슬릿 집속 장치섬광검출기 배열이중모드 신호 처리 모듈
제목
다중 슬릿 즉발감마선 카메라를 위한 이중모드 신호처리 모듈 개발
제목 (타언어)
Development of Dual-mode Signal Processing Module for Multi-slit Prompt-gamma Camera
저자
박종훈신동호이세병이한림김찬형김성훈정종휘
DOI
10.14316/pmp.2016.27.1.37
발행일
2016-03
저널명
PROGRESS in MEDICAL PHYSICS
27
1
페이지
37 ~ 45