Carbon dioxide measurement from breath gas using blind source separation

Citations

SCOPUS

0

초록

There are many approaches which use breath gas to diagnose disease such as lung cancer, breast cancer and diabetes. But, it has not been used in medical field because of low accuracy by low concentration of target component (ppt, ppb level). According to previous study, end-tidal concentrations of Volatile Organic Component (VOC) correlated with carbon dioxide (CO2) levels. So, we propose CO2 measurement method from breath gas using blind source separation. To verify feasibility of my approach, we measured 132 samples from 10 persons (Age 22-36, 6 males, 4 females) and calculate standard deviation of measured signal and source signal. At the 0.05 level, the difference of the standard deviations is significantly different (p=0.001). According to this result, it shows our approach is applicable to measure CO2 from mixed breath gas. And, it enables to measure VOC using breath gas.

키워드

Breath gasCarbon dioxideComponentVolatile organic componentApartment housesBiocommunicationsBioinformaticsBiomedical engineeringCarbon dioxideConcentration (process)Information scienceIsomersOrganic carbonOxidesSeparationSignal analysisStatisticsVolatile organic compoundsBreath gasComponentInternational conferencesSource separationVolatile organic componentBlind source separation
제목
Carbon dioxide measurement from breath gas using blind source separation
저자
Lee, JongyounBai, JingKim, In Young
DOI
10.1109/ICBBE.2008.849
발행일
2008-05
유형
Conference Paper
저널명
2nd International Conference on Bioinformatics and Biomedical Engineering, iCBBE 2008
페이지
2080 ~ 2082