Estimation of Arterial Radial Wall Displacement Using a Low-Profile Ultrasonic Transducer

Citations

SCOPUS

0

초록

Blood pressure is a vital indicator for cardiovascular diagnosis, and its continuous monitoring requires wearable devices. This study proposes a Doppler-based method to estimate radial arterial wall displacement, validated through Field II simulations and in vivo measurements using wearable CMUT and PZT-4 transducers. RF data were acquired at 50 ms intervals with a 62.5 MHz sampling rate, and displacement was estimated via the Loupas method. Results showed arterial diameters of 2.3–2.4 mm and wall displacements of 30–50 µm, consistent with reported values. The waveform reproduced systolic and diastolic phases as well as the dicrotic notch, while beam profile effects on estimation accuracy were limited.

키워드

Blood pressure monitoringradial arterywall displacementDoppler ultrasoundLoupas methodCMUTPZTBloodDiagnosisDoppler effectSignal processingSolid-state sensorsWearable technology
제목
Estimation of Arterial Radial Wall Displacement Using a Low-Profile Ultrasonic Transducer
저자
Kim, Jin-hyukKim, JeonghoonKim, HyunsuPark, Kwan Kyu
DOI
10.1109/IUS62464.2025.11201798
발행일
2025-10
유형
Conference paper
저널명
IEEE International Ultrasonics Symposium, IUS
페이지
1 ~ 4