상세 보기
RF Heating of Implants in MRI: Electromagnetic Analysis and Solutions
RF Heating of Implants in MRI: Electromagnetic Analysis and Solutions
- Cho, Youngdae;
- Yoo, Hyoung suk
초록
When a patient takes an MRI scan, the patient has a risk of unexpected injuries due to the intensive electromagnetic (EM) field. Among the injuries, the tissue heating by the time-varying EM field is one of the main issues. Since an implanted artificial structure with a conductive material aggravates the heating effect, lots of studies have been conducted to investigate the effect around the implants. In this review article, a mechanism of RF heating around the implants and related studies are comprehensively investigated.
키워드
Radio Frequency; Heating; Implants; RF coils
- 제목
- RF Heating of Implants in MRI: Electromagnetic Analysis and Solutions
- 제목 (타언어)
- RF Heating of Implants in MRI: Electromagnetic Analysis and Solutions
- 저자
- Cho, Youngdae; Yoo, Hyoung suk
- 발행일
- 2020-00
- 권
- 24
- 호
- 2
- 페이지
- 67 ~ 75