Graphene nanoplatelet/graphitized nanodiamond-based nanocomposite for mediator-free electrochemical sensing of urea
- Kumar, Vanish;
- Kaur, Inderpreet;
- Arora, Saloni;
- Mehla, Ravi;
- Vellingiri, Kowsalya;
- ... Kim, Ki-Hyun
WEB OF SCIENCE
52SCOPUS
63초록
Urea is well-known to offer tremendous scope for sensing/diagnosing such as adulteration in dairy products or diseases in human body. This study was organized to describe and validate a new mediator-free, unsophisticated, and direct current voltage (IV)-based sensor for facile detection of urea using nanocomposites made of urease-immobilized graphene nanoplatelets and graphitized nanodiamonds. This nanocomposite displayed sensitive and direct signal in the form of current at 0 V without the need of any complex chemical reaction. This platform was highly sensitive (limit of detection of 5 mu g/mL) far superior to the comparable systems introduced recently. The incorporation of graphitized nanodiamonds within the graphene nanoplatelets layers helped improve the sensitivity by a factor of three (up to 806.3 mu A (mg mL(-1))(-1) cm(-2)) with 20 s response time. As such, the use of this nanocomposite was helpful in improving sensing performances with enhanced enzyme loading capacity.
키워드
- 제목
- Graphene nanoplatelet/graphitized nanodiamond-based nanocomposite for mediator-free electrochemical sensing of urea
- 저자
- Kumar, Vanish; Kaur, Inderpreet; Arora, Saloni; Mehla, Ravi; Vellingiri, Kowsalya; Kim, Ki-Hyun
- 발행일
- 2020-01
- 유형
- Article
- 저널명
- Food Chemistry
- 권
- 303
- 페이지
- 1 ~ 8