Estimation of blood pressure using photoplethysmography on the wrist

  • Song, Soo Hwa
  • Cho, J.S.
  • Oh, Hong Sic
  • Lee, Jong Shill
  • Kim, In Young
Citations

SCOPUS

22

초록

Blood pressure measuring method using PPG signal is the one of many non-invasive blood pressure methods. In this study, we have studied a new method that estimates the blood pressure using wrist cuff and wrist PPG (Photoplethysmograph) signal. During the deflation of the wrist cuff pressure, the PPG pulse appears at certain point that is similar to Korotkoff sound. After the pulse appeared, the morphology of PPG pulses was changed into the certain shape. So, we use these points to estimate the blood pressure. For validation, the reference SBP and DBP are measured by auscultatory method at the left upper arm. Subjects were 13 non cardiac disease of man in their twenties. The result of this study, the mean difference of SBP and DBP, between auscultatory method and PPG analysis is 2.08 ± 5.07mmHg and 1.38 ± 7.86mmHg respectively.

키워드

Cardiac diseaseCuff pressureNon-invasivePhotoplethysmographPhotoplethysmographyUpper armBloodCardiologyBlood pressure
제목
Estimation of blood pressure using photoplethysmography on the wrist
저자
Song, Soo HwaCho, J.S.Oh, Hong SicLee, Jong ShillKim, In Young
발행일
2009-09
유형
Conference Paper
저널명
Computers in Cardiology
36
페이지
741 ~ 744