Validation of intra- and interobserver reproducibility of shearwave elastography: Phantom study

  • Mun, Han Song
  • Choi, Seon Hyeong
  • Kook, Shin Ho
  • Choi, Yoonjung
  • Jeong, Woo Kyoung
  • ... Kim, Yongsoo
Citations

WEB OF SCIENCE

36
Citations

SCOPUS

36

초록

Purpose: To assess intra- and interobserver reproducibility of quantitative ultrasound (US) elastography for breast masses using shearwave elastography (SWE). Materials and methods: The US elastography phantom was used. This training phantom contained 12 masses with dimensions from 6 to 11 mm and varying B-mode (hypo- iso- and hyperechoic) characteristics and viscoelastic properties varying from about 50 to 250 kPa. Four attending radiologists participated as operators and used Supersonic Imaging 15-4 MHz array to acquire images. In each session, B-mode US features were assessed once and US elastography values were assessed twice in a consecutive and distinct manner. A total of three identical sessions of data acquisition were performed at 2-week intervals. For categorical variables such as BI-RADS features of mass on B-mode image and qualitative SWE features, kappa (k) values were calculated by the generalized linear mixed model. For continuous variables, kPa, and ratio, intraclass correlation coefficients (ICCs) were calculated. Results: Interobserver agreement for SWE image similarity was moderate (k = 0.57). With respect to quantitative SWE values, intraobserver ICCs were 0.65, 0.77, 0.92, and 0.91 for maximum elasticity; 0.70, 0.83, 0.94, and 0.94 for mean elasticity; and 0.67, 0.83, 0.92, and 0.92 for elasticity ratio for operators 1, 2, 3, and 4, respectively. Interobserver reproducibility showed good agreement with ICC values of 0.77 for maximum elasticity, 0.82 for mean elasticity, and 0.79 for elasticity ratio. Conclusion: Quantitative US elastography for breast masses was highly reproducible with good agreement across operators. With respect to intraobserver reproducibility, ICC values varied but three out of four operators showed good agreement. (C) 2013 Elsevier B. V. All rights reserved.

키워드

UltrasoundElastographyShearwave elastographyReproducibilityPhantomBREAST-LESIONSMASSES
제목
Validation of intra- and interobserver reproducibility of shearwave elastography: Phantom study
저자
Mun, Han SongChoi, Seon HyeongKook, Shin HoChoi, YoonjungJeong, Woo KyoungKim, Yongsoo
DOI
10.1016/j.ultras.2013.01.013
발행일
2013-07
유형
Article
저널명
Ultrasonics
53
5
페이지
1039 ~ 1043