Measurement of electron temperature and ion density using the self-bias effect in plasmas

Citations

WEB OF SCIENCE

25
Citations

SCOPUS

24

초록

For novel plasma diagnostics, the rf floating probe was revisited. For inducing the self-bias effect, ac bias voltage (similar to kilohertz) was applied through a dc blocking capacitor between a probe and a signal generator. The dc self-bias potential was changed not only with ac bias voltages but also with electron temperatures, and therefore, the electron temperature was derived from the variations in the self-bias potential with and without ac bias voltage. The harmonic component of the probe contains information about the ion flux, and using a fast Fourier transform analysis of the probe current, the ion density was derived from the first harmonic current of the probe. The experimental results were compared with a single Langmuir probe. The electron temperature and the ion density were in good agreement with those from the Langmuir probe. Because the amplitude of the ac bias voltage is very low (<3 V), local ionizations affected by a high bias-voltage can be neglected.

키워드

ENERGY DISTRIBUTION
제목
Measurement of electron temperature and ion density using the self-bias effect in plasmas
저자
Hwang, Kwang-TaeOh, Se-JinChoi, Ik-JinChung, Chin-Wook
DOI
10.1063/1.3430632
발행일
2010-06
유형
Article
저널명
Physics of Plasmas
17
6
페이지
1 ~ 4